GE AKTA avant User Manual page 329

Hide thumbs Also See for AKTA avant:
Table of Contents

Advertisement

Step
10
11
12
13
14
ÄKTA avant User Manual 29035184 AE
Action
In the Run Data pane of System Control, read the current Cond temp.
Tip:
If Cond temp is not showing, click the Customize icon. In the Customize dialog
box, under the Run Data Groups tab, select to view Cond temp.
In the Calibration dialog box, enter the theoretical conductivity value at the
current conductivity temperature in the Enter theoretical conductivity
value input field.
If a certified conductivity standard solution is used, use the supplied
theoretical conductivity value.
If a manually prepared 1.00 M NaCl calibration solution is used, see the
graph for conductivity value at the current temperature Graph for con-
ductivity value, on page 330.
In the Calibration dialog box, click Calibrate.
Result: The new conductivity cell constant is displayed in the Conductivity
cell 1 constant/cm box. The new constant should normally be 40 ± 10 cm
The date and time for the calibration are also displayed.
In the System Control toolbar, click the End icon to end the run.
In the System Settings dialog box, select Conductivity:Cond temp compen-
sation and set the Compensation factor back to desired value, default 2.1%.
Click OK.
7 Maintenance
7.7 Calibration procedures
7.7.3 Calibrate the Conductivity Monitor
-1
.
329

Advertisement

Table of Contents
loading

Table of Contents