5 Operation
5.8 Fractionation
5.8.2 Prepare Fraction collector F9-R
Step
3
4
5
216
Action
Adjust the height so that the horizontal mark (5) on the tube sensor (6)
•
is at the same level as the top of the flat collection tubes and approxi-
mately 2 mm over the top of the flanged collection tubes (4).
Lock the delivery arm at this height with the lock knob.
•
Check that the tube sensor (9) is in the correct position for the tubes used
(10). The eluent tubing (7) should be above the center of the collection tube.
7
8
9
10
Use the sensor control (8) to position the tube holder over the center of the
collection tube.
4
6
5
ÄKTA avant User Manual 29035184 AE