Purge System Pump A
Purge both pump heads of System Pump A by following the same procedure as in Purge
System Pump B, on page 168, but replace step 3 and 4 with the following:
Step
1
2
ÄKTA avant User Manual 29035184 AE
Action
In the Process Picture:
Click the Inlet valve A icon.
Click the position of one of the inlets that will be used at the beginning
of the run.
Result: The inlet valve switches to the selected port.
In the Process Picture:
Click the System pumps icon.
Set Conc % B to 0% B and click Set % B.
Result: Only System Pump A is active.
5.4 Prime inlets and purge pump heads
5.4.1 Prime buffer inlets and purge system pumps
5 Operation
171