Troubleshooting flowchart

Start
Open Internet Explorer and type:
https://198.169.188.10:443
Returns "GE
Healthcare
Web Server"?
Re-check configuration and reboot.
System should work.
7 - 176
PROPRIETARY TO GE
. 7
EV
Figure 7-144 Troubleshooting flowchart
No
Yes
Section 7-7 - Troubleshooting Insite ExC Configuration
LOGIQ E9 P
The hospital is not allowing the system to route out.
Request IT to open a path to the Internet port 443
(Secure Socket Layer).
Request IT to provide internet access.
Message
returned?
(Access
Denied)
The hospital firewall is blocking your access to the
Internet. Request IT to allow access.
Also, check is a proxy server is needed.
Request IT to provide proxy information.