Structured Reporting

Sending Structure Report - R3.x and earlier

Sending Structure Report - R4.x and later

A Structured Report can be sent for the measurements and calculations of the currently-open exam
via a Print button. Add the Image Storage service to the Printflow setting of that button and set the
Format for that button to "M&A." Take measurements in an exam and press the Print button. The
LOGIQ E9 generates a Structured Report and sends it to the Image Storage service.
The LOGIQ E9 does not store Structured Report files. Instead, when called upon, it conveys the
measurements and calculations to the service via the DICOM message exchange between them.