Utility -> Connectivity -> Device Screen - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
3-9-1
Setup the LOGIQ E9 for DICOM Communications (cont'd)
Table 3-8
1.
2.
3.
PROPRIETARY TO GE
. 7
EV
Utility -> Connectivity -> Device screen
Select MyComputer.
Assign an AE Title to the LOGIQ E9. (AE stands for Application Entity. DICOM services use
this to identify the LOGIQ E9.) AE Title is case-sensitive. This title may contain the Computer
Name from the TCP/IP page, if desired. Maximum number of characters in AE Title is 16
characters.
NOTE: It is NOT recommended to use the factory default. This is not prohibited, but more than
one system with the same AE Title can cause confusion.
Edit the Port Number if needed. "104" is typical. Save your changes and reboot the system.
Chapter 3 LOGIQ E9 Setup
LOGIQ E9 P
ROPRIETARY
S
M
ERVICE
ANUAL
3 - 41

Advertisement

Table of Contents
loading

Table of Contents