Op Panel Utilities - Op Panel Interface; Accessing The Service Platform From Windows - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-5-10

OP Panel Utilities - Op Panel Interface

The Operator Panel functionality can be tested using a program (GE Test App) available through the
Service Platform.
R4 or earlier
R5 and later, OP Panel Test only available on Windows desktop.
Although this diagnostic may be executed when the application software is running, its performance is
greatly reduced and the controls will respond very slowly. It is recommended that this test be run from
Windows before the application software has started. To access this test from Windows, see the
instructions below.
7-5-10-1

Accessing the Service Platform from Windows

1.) Insert the Service Key into a USB port.
2.) Press the Standby (ON/OFF) button and select Exit.
NOTE:
If the system exit menu does not appear, press Alt F10.
With the service dongle attached, a screen will be displayed to allow Maintenance Access.
3.) In the Maintenance Access screen, leave the Username blank, enter the current Password and
select OK.
4.) From the Start Application screen select Maintenance.
5.) In the Maintenance window select Exit to Windows to access the Windows Desktop.
6.) Click on the GETestApp for the operator panel test.
PROPRIETARY TO GE
. 7
EV
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
ROPRIETARY
S
M
ERVICE
ANUAL
7 - 45