Essential Tests; Essential Tests Differences In R4 And Later; Essential Tests Window - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-5-14-1

Essential Tests

See:
7-5-7 "Diagnostics Window Overview" on page 7-30
Runs tests on essential functions of the BEP. This is a good starting point for isolating issues that may
originate from the BEP. The test results are displayed in the Status portion of the window.
Run Time: About 2 minutes.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button.
3.) Failed tests may have another diagnostic test that can help isolate the failure, or will indicate which
part to replace. For:
-
-
-
-
-
-
7-5-14-2

Essential Tests differences in R4 and later

Essential test in R4 has two versions:
Essential test: Complete test, same as R3 or earlier
Essential test, No network: same test, except it does not run the network tests for wired or wireless
connection.
PROPRIETARY TO GE
. 7
EV
PCI Bus configuration
CMOS memory
CPU - run
7-5-14-3 "System Board" on page 7-115
RAM - run
7-5-14-12 "Memory" on page 7-123
Hard Drive status and Random Seek - run
and run
7-116
7-5-14-5 "Hard Disk Drive Quick Test" on page 7-117
Network Interface (loopback only) - run
Figure 7-78 Essential Tests Window
When network cable or Wireless network does not exist, the
Essential test fails but does not translate into a hardware
failure.
Essential test No Network will not run Network tests to quickly
identify other failures not related to network. (Intended for use
mostly in Manufacturing)
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
ROPRIETARY
for a description of this window's features.
7-5-14-4 "Hard Disk Drive Surface Scan" on page
7-5-14-10 "Network Adapter" on page
S
M
ERVICE
ANUAL
7-121.
7 - 113