I/O Board Tests; I/O Board Tests Window - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-5-12-11

I/O Board Tests

See:
7-5-7 "Diagnostics Window Overview" on page 7-30
features.
Symptoms:
Intermittent Problems
Popup messages warning of system voltage problems - I/O Board Pwr Supply Test
System comes up in simulator mode - run I/O Board GFI Voltage Test
Unable to run diagnostics
No doppler audio/system audio/general audio sounds - run I/O Board Audio Test
No display on the monitor
1.) Select the I/O Board Tests checkbox to run all the subtests, or select the individual subtests to run
only the selected tests.
2.) Enter the desired Loop Count (numeric value of 4 digits or less).
3.) Click Execute to run the test(s).
PROPRIETARY TO GE
. 7
EV
Figure 7-60 I/O Board Tests Window
Chapter 7 Diagnostics/Troubleshooting
LOGIQ E9 P
ROPRIETARY
for a description of the looping and list
S
M
ERVICE
ANUAL
7 - 83