S-Video Image Displays A Portion Of The Main Display; The Monitor Measurement Summary Sidebar Does Not Appear After Upgrading To R6 - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
7-5-14-7

S-Video Image displays a portion of the Main Display

The S-Video image displays a portion of the main display that includes the probe image in both single
and dual probe display modes. To provide increased resolution and maintain the proper aspect ratio on
the S-Video display, the entire main display image is not projected on the S-Video monitor. The S-Video
crop area was selected to optimize the probe image so when the LOGIQ E9 is not in an imaging mode,
the S-Video will continue to display a cropped portion of the screen which may appear incorrect.
7-5-14-8

The Monitor Measurement Summary Sidebar does not appear after Upgrading to R6

After upgrading to R6, the monitor Measurement Summary Sidebar does not appear, and the image
uses the same space it would for Dual, all the time. So the image will go right up to the left edge of the
screen.
When upgrading to R6 software in a LOGIQ E9 with a 19 inch LCD, the following steps must be
completed to have the correct layout for 19 inch Monitors. DO NOT skip this step, or the monitor layout
will not be correct.
Go
-
-
PROPRIETARY TO GE
. 7
EV
o Utilities -> System -> System Display:
t
Set the drop-down menu in Use Wide Screen for... > Single Image to OFF
Make sure the checkbox in Side Panel Content > Measurement Summary is checked.