Structured Reporting - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
3-9-10

Structured Reporting

Structured Reporting provides for transmission of exam measurements and calculations in a separate
non-image file. It is not a service in itself, but is a feature of some DICOM Image Storage services, and
is always associated with one. The service must be able to interpret and display the data in the
Structured Report. Refer to the documentation of the Image Storage service documentation for
Structured Reporting.
Table 3-26
To enable Structured Reporting for an Image Storage service, go to the set-up screen for that service
in Utility -> Connectivity -> Service, check the Enable Structured Reporting box and press Save.
From that point on, the LOGIQ E9 sends a Structured Report at the end of an exam with
measurements.
PROPRIETARY TO GE
. 7
EV
Structured Reporting
Utility -> Connectivity -> Service - Enable Structured Reporting checked
Chapter 3 LOGIQ E9 Setup
LOGIQ E9 P
(continued on next page)
S
M
ROPRIETARY
ERVICE
ANUAL
3 - 73

Advertisement

Table of Contents
loading

Table of Contents