Touch Panel

7-5-17-1

Calibration

NOTE:
Always try TP calibration verification before attempting to calibrate.
Verification" on page
calibrate and the normal calibration procedure does not work, it may be necessary to do a
complete Pre-calibration. See:
To do a Touch Screen Calibration, follow the directions on the Touch Screen. As each of the cross-hairs
appear, touch them with your finger or a pencil eraser.
NOTE:
You MUST hold your finger on the cross-hair until it moves to the next location. If you just tap
the calibration cross-hair, there is a good chance your calibration will be corrupt.
7 - 130
PROPRIETARY TO GE
. 7
EV
7-131. If the Touch Panel is so far out of calibration, hangs or does not
Section 6-5 "Touch Panel Calibration" on page
Figure 7-92 Touch Panel Calibration Screen
Section 7-5 - Common Service Desktop
LOGIQ E9 P
ROPRIETARY
7-5-17-2 "Calibration
6-19.
S
M
ERVICE
ANUAL