Configuration Steps (cont'd)
Figure 3-31 Service Login Screen - LOGIQ E9
4.) Select GE Service and enter the password, then select OKAY.
The Service Desktop opens.
5.) Select the CONFIGURATION tab.
The Configuration Page opens (Figure 3-32 "Configuration Home Page - LOGIQ E9" on page 3-88).
3-11-4

Configuring InSite ExC

Configure the System to connect to InSite ExC. All bold fields are mandatory. Before configuring the
agent be sure that you have a network connection and check that it is configured. You will not be able
to perform checkout with the default network configuration.
3-11-4-1

Configuring Agent Configuration

1.) Select the CONFIGURATION tab and choose the InSite ExC Agent Configuration link in the
frame on the left side.
2.) Inspect the Device Name field and the CRM field. The Device Name field auto populates with the
prefix LE9_ followed by the Serial number.
NOTE:
The Device Name cannot be edited.
3.) In the Serial Number/System ID (CRM) field (for Ultrasound), enter the System ID now.
This is a required field. For consoles located in the U.S.A., the System ID is pre-populated. Outside
of the U.S.A., follow the local System ID convention.
The CRM field auto-populates with format of LOGIQ E9 followed with the first five digits of the serial
number. The CRM field is editable, and can be edited to reflect the desired System ID.
NOTE:
If CRM says unknown, try rebooting. When you install software the system reboots, but you
need a second boot to get the serial number into the service platform.
PROPRIETARY TO GE
. 7
EV
Chapter 3 LOGIQ E9 Setup
LOGIQ E9 P
ROPRIETARY
S
M
ERVICE
ANUAL
3 - 87