General Requirements; Electrical Specifications For The Logiq E9; Inrush Current; Electrical Requirements For Logiq E9 - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
2-2-3
Electrical requirements
2-2-3-1

General requirements

NOTE:
GE requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This
dedicated power shall originate at the last distribution panel before the LOGIQ E9.
The LOGIQ E9 will function on Voltages from 100-240 Volts and 50 or 60 Hz. However, if using 220 volt
power, then a center tapped power source is required.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the LOGIQ E9 outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the LOGIQ E9 outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the ground from the
main facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-3-2

Electrical requirements for LOGIQ E9

In the table below, the electrical specifications for LOGIQ E9 include LCD and on board peripherals.
Table 2-3
PART NUMBER
DESCRIPTION
5205000-x
LOGIQ E9, 100-240 VAC
The current drain varies, depending on the mains voltage.
At 230 VAC, the current may be up to 8 A.
At 100 VAC, the current may be up to 10 A.
2-2-3-3

Inrush current

During power on, an inrush circuit prevents the current from increasing above the stated values.
Table 2-4
VOLTAGE
90 VAC
110 VAC
220 VAC
264 VAC
PROPRIETARY TO GE
. 7
EV

Electrical specifications for the LOGIQ E9

VOLTAGE
100-240 VAC
Inrush current at different mains voltages
50 Hz
13 A
9 A
5.5 A
6 A
Chapter 2 Site preparations
LOGIQ E9 P
POWER
TOLERANCES
CONSUMPTION
+/-10%
1100 W
60 Hz
12 A
11 A
6 A
5 A
S
M
ROPRIETARY
ERVICE
ANUAL
FREQUENCY
50/60 Hz
2 - 3