PROPRIETARY TO GE
D
5180483-100, R
. 7
LOGIQ E9 P
S
M
IRECTION
EV
ROPRIETARY
ERVICE
ANUAL
3-9-7
DICOM Worklist Service (cont'd)
Table 3-19
New Field in Worklist Setup
R5.1.1 and earlier
R5.1.2 and later
Verify Before Query: If the LOGIQ E9 attempts a Query to a service that does not respond, the
scanner will wait for a period determined by the Timeout and Maximum Retries settings. During this
period user input is disabled. To avoid this, checkmark Verify Before Query. Then the scanner will first
verify that the service will respond, and wait no longer than "Verify Timeout" seconds. If the service
fails to respond in that time, the LOGIQ E9 will not attempt the Query.
3 - 64
Section 3-9 - Connectivity Setup and Tips