Required Facility Needs - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
2-3-3

Required facility needs

NOTE:
GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the system.
The LOGIQ E9 will function on Voltages from 100-240 Volts and 50 or 60 Hz. However, if using 220 volt
power, then a center tapped power source is required.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Dedicated single branch power outlet of adequate amperage (see:
specifications for the LOGIQ E9" on page
less than 2.5 m (8 ft.) from the LOGIQ E9's proposed location
Door opening is at least 76 cm (30 in) wide.
Proposed location for LOGIQ E9 is at least 0.46 m (18 inches) from the wall or objects for cooling.
Power outlet and place for any external peripheral are within 2 m (6.5 ft.) of each other with
peripheral within 1 m (3.2 ft.) of the LOGIQ E9 to connect cables.
Power outlets for other medical equipment and gel warmer.
Power outlets for test equipment within 1 m (3.2 ft.) of LOGIQ E9.
Clean and protected space to store transducers (in their cases or on a rack).
Material to safely clean probes (done with a plastic container, never metal).
2 - 8
PROPRIETARY TO GE
. 7
EV
Section 2-3 - Facility needs
LOGIQ E9 P
2-3) meeting all local and national codes which is located
S
M
ROPRIETARY
ERVICE
Table 2-3 "Electrical
ANUAL