Spooler (cont'd)
Common Error Messages
No connection established – check network and log file: This usually indicates a bad
or missing network connection OR incorrect IP Address, AE Title, or Port # between the LOGIQ E9 and
service.
The remote system was not able to interpret the image information – check log
file: This indicates an issue with the image or patient data, such as a missing Patient ID.
The remote service is out of resources: For example, the PACS has no more storage space.
3-9-13

Wireless

Configuring a Wireless network is beyond the scope of this guide. Both the User and Service manuals
cover Wireless setup in detail. In particular, for setting up the WLAN, refer to the appropriat e version the
LOGIQ E9 Basic User Manual, Chapter 16, or the appropriate LOGIQ E9 Release Notes. See:
8 "Basic User Manual and Release Notes per LOGIQ E9 console" on page
Wireless Network (Option). (Available via the "Help" F1 key on the LOGIQ E9 keyboard.)
Table 3-31
NOTE: When both wired and wireless are enabled, the LOGIQ E9 chooses wired.