Configuration Home Page - Logiq E9 - GE LOGIQ E9 Service Manual

Hide thumbs Also See for LOGIQ E9:
Table of Contents

Advertisement

D
5180483-100, R
IRECTION
3-11-4-2
Configuring Advanced Configuration (cont'd)
7.) Access and inspect the CSD Home Page. The Home Page appears and indicates Configured and
Checked out.
NOTE:
If this fails wait a minute and refresh the page by clicking on the home page button again. There
is often a delay before the checkout is complete.
Figure 3-35 Configuration Home Page - LOGIQ E9
PROPRIETARY TO GE
. 7
EV
If configured properly, the Service Home Page updates in approximately 10 seconds
(although it may take up to 2 minutes). If it does not check out, confirm the Agent Configuration
Tool screen has correct URL / IP address information. Make corrections, and repeat 1 Select
PILOT2 or PRODUCTION from the Enterprise Server dropdown list. The Pilot server is used
for initial production, please check with the OLC to verify that the current server is used. 3-50.
If it fails again, contact the OLC or the Hospital IT department to check that your system has
access to the Internet.
Chapter 3 LOGIQ E9 Setup
LOGIQ E9 P
ROPRIETARY
S
M
ERVICE
ANUAL
3 - 91

Advertisement

Table of Contents
loading