Mouse (Trackball)
See:
for a description of this window's features.
7-5-7 "Diagnostics Window Overview" on page 7-30
Runs a test on essential functions of the Trackball. The test results are displayed in the Status portion
of the window.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button.
3.) Verify the signal cable is connected securely.
4.) Clean the Trackball. If the problem continues:
5.) Replace the Trackball.
Figure 7-90 Mouse Window