How FE Remotely Enables Disruptive Mode and VCO (cont'd)
6.) The customer selects Yes.
NOTE:
If the customer does not wish to have diagnostics running at the time of the request, they select
No. A message is sent back to the OLC or FE that "Disruptive Mode" is not enabled.
The screen below confirms the Disruptive Mode status change.
7.) After "Disruptive Mode" is enabled, "Virtual Console Observation" can be activated by selecting
Utilities -> Common Utilities -> Virtual Console Observation, and then pressing the "Start"
button.
If "Virtual Console Observation" has been activated successfully, the following display is shown to the
OLC or FE:
8.) To terminate "Virtual Console Observation", the OLC or FE can press the "Stop" button.
PROPRIETARY TO GE
. 7
EV
Figure 7-137 Disruptive Mode Enabled
Figure 7-138 Virtual Console Observation is Running
Chapter 7 Diagnostics/Troubleshooting