The SETUP Program
9.5 LIMITS
to the main menu.
Limit fest
1 O%eoHY
[or NJ
The program asks if a limit test is required.
Press
to request a limit test, or
to begin entering the
next group.
After the limit type has been specified, proceed to subsection 9.3. 3, "Limit Actions."
9.5.1 Analog
Limit typed Low High
or
Limit
point?
limit point
ENDLINE
The program asks you to specify either a lower limit
(Lid) or an upper limit (Hi ah). The limit action will be
triggered when any measurement in the channel range
being scanned goes past the specified limit point. (The
action is triggered when a measurement falls below a
lower limit, or when a measurement rises above an upper
limit.)
Press
to select a low limit or
to select a high limit. If
average has been specified, the limit test is done on the
average, not the individual readings.
The program prompts for the value to be compared with
each reading.
Enter the limit point, and press ENDLINE | .