System Monitoring
The limit action menu appears. Just press
to
select EEEF as the action. The HP-71 will beep if the
temperature either rises above 26 °C or falls below 24°C.
Note that if the beep is not a satisfactory alarm, you could
use an actuator channel to trigger an external alarm bell
or other signal.
Now you can set the control (actuator) limits.
To set a lower limit, press [L].
Set the lower limit to 24. 5°C. The program will compare
each reading with this limit.
The limit action menu appears. Use the
key to select
CLLOSE ACT-BIT asthe action.
Select channel 0 as the actuator channel. Whenever the
temperature of the bath drops below 24.5°C, the HP-71
will close actuator channel O on the HP 3421A, which
actuates the heater relay.
Now you can set the upper control limit.
To set an upper limit, press [H].
Set the upper limit to 25. 5°C.