Software Procedures
B
NOTE: All settings need to be set individually for each monitor. In
addition to this resolution setting, monitor 1 should be set as Primary
Monitor and Extend Windows Desktop box should be checked, monitor 2
& 3 & 4 (if applicable) should be attached and Extend Windows Desktop
box should be checked
Table 6-1: RA600 Workstations Resolution and Frequency Settings
Note:
You might have to execute the following test a second time if the monitor was blank initially.
12.
Click the Test button and then click OK. Check that the Hi-Res monitors display the test
pattern.
Note:
A test pattern should be displayed on monitor one. If not, check the settings, connections, and
power. Repeat the test until a test pattern is displayed. Click on Yes.
13.
Click OK to make the changes permanent and then click Yes to restart the workstation.
14.
Reboot the Workstation, re-login as Administrator and return to the Display Panel.
15.
Select the DOME tab from the Display Panel.
16.
Configure the Panel as listed below:
a.
Select Dynamic Gray (from the palette option drop box).
b.
Select the appropriate resolution that matches that as configured in the Settings table,
above.
c.
Select the correct monitor refresh rate that matches that as configured In the Settings table,
above.
MD2 PCI
1200 x 1600
See Note
Page 195
SW 004 (A+,A,B) Dome Video Board Driver Installation
75.5 Hz