Finally, when the displays are calibrated again, retry the tasks with poor results. (See section
"Run an individual task" for more information about executing a single task.)
5-2 Measure Uniformity Task is 'WARNING' or 'NOT OK'
1.
In case a result of the Measure Uniformity task is unsatisfying, find out which display is causing
the problem. See section "Look up histories of executed tasks" if you need more information
about retrieving detailed information on task results.
2.
Once you know which display is responsible for the problem, you should check if the LUC
option (Luminance Uniformity Correction) of that display is enabled. Follow the instructions
below to find the LUC setting.
CNF 006 Monitor Calibration with Barco Tool (MediCal Pro)
Page 124