Functional Checks / Setup Completion - GE Centricity RA600 Service Manual

Table of Contents

Advertisement

Functional Check
n
VF 001

Functional Checks / Setup Completion

Section 1: Introduction
This job card finishes the required system configurations before turnover to the customer. This
procedure will:
u
Test to verify valid network connections.
u
Set the administrator password.
u
Launch the application program and check functionality
Section 2: Prerequisites
u
Workstation integration completed.
u
Workstation operating system software installed and configured.
u
Workstation application software installed and configured.
u
Monitors calibrated.
Section 3: Procedure
1.
Login as administrator.
No password is set at this time. If a password has been set it could be: (CANal1).
2.
Press the Enter key.
Page 162