Software Configuration
7.
Close the License Configuration window.
3-3 Install Application Software Patch
Examine the installation package for any software update patch CD for the RA600 program. When
any such patch is included, install the patch at this point per job card
3-4 Initialize MediCal Pro Software
1.
Launch MediCal Pro software by clicking Start > Programs > MediCal Pro > MediCal Pro.
2.
Enter the license code when prompted.
The license code can be found inside the MediCal Pro package, on the back of the CD case.
3.
Set up and configure the MediCal Pro software according to job card
Calibration with Barco Tool (MediCal
Note:
The MediCal Pro software is not intended for use with C monitors. The software can be used to
check C monitor calibration but since C monitor brightness and contrast controls are available
on the front of the monitor, no assurance is possible that the settings will remain (manually)
unchanged between calibrations
3-5 Initialize Centricity RA600 Software
Launch the RA600 software by double clicking the icon next to the Start button or by selecting
Start > Programs > Centricity RA600 > Centricity RA600.
CNF 001 Initial Power Up and Software Configuration