2.4

WIRING THE RECORDER

2.4.1 Recommendations
All wiring must be in accordance with local electrical codes and should be carried out by
authorized experienced personnel.
The ground terminal must be connected before any other wiring (and disconnected
last).
A switch in the main supply is recommended near the equipment.
If an external fuse is used to protect the electrical circuit to the recorder, the fuse
should match the recorder fuse rating (fuse type) as well as for the fuse holder.
Sensor wiring should be run as far as possible from power wiring. (motors, contactors,
alarms, etc.)
To reduce stray pick-up, we recommend the use of a twisted pair sensor wiring.
EMI effects can be further reduced by the use of shielded cable sensor wiring. The
shield must be connected to the ground terminal.
The use of spade terminals on all wiring is recommended.
2-7
2. INSTALLATION