Content Of The Data Register; Access Rights; Main Tab (Main Display) - Honeywell EK205-M Operating Instructions Manual

Volume conversion device
Table of Contents

Advertisement

Operation
9.4

Content of the data register

9.4.1 Access rights

The "Access" column in the tables of the following sections describes what lock needs to be
opened to change a parameter. All changes of parameters are saved in a logbook.
Access
Meaning
C
Calibration officer
15
CDL
Calibration Manual
A
Administrator
Cu
Customer
-
No access

9.4.2 Main tab (main display)

In the Main tab, the key measurement values are displayed at a glance. The content can differ
depending on the settings (see section 9.4.6.5). With this tab, there is no possibility to move the
16
cursor for a value.
Via the ENTER key, an update of the values displayed can be initiated.
Display
Meaning
Vb
Volume at base conditions
VmT
Total Volume at measurement conditions
p
Pressure
T
Temperature
9.4.2.1 Vb – volume in base mode
The measured volume of the gas meter connected is converted in base mode and added up in
this counter.
If a fault is given (see section 11 Faults), the counter is stopped and the volume is added up in a
special fault volume counter.
9.4.2.2 VmT – Total volume in operation mode
The total measured volume of the gas meter connected is added up in this counter.
Even if a fault is given (see section 11 Faults), the counter is still counting. Additionally, the
disturbed volume is counted in a special fault volume counter VmD.
15
If a calibration logbook is available in the device.
16
An exception is if the Main tab was set to List (see section 9.4.6.5).
Unit
Access
3
m
C
3
m
C
bar
-
°C
-
91
Address
2:300
4:302
7:310_1