Honeywell EK205-M Operating Instructions Manual page 73

Volume conversion device
Table of Contents

Advertisement

Data transfer
8.2.1.2 EK205-M parameterization
i
To parameterize the EK205-M, you will require a connection cable for the optical
interface as well as the enSuite program, which can be downloaded free of charge
from
www.elster-
a) Closing the time windows
Time windows 3 and 4 must be closed for parameterization to function securely. To do this:
Connect a PC or laptop with enSuite installed to the EK205-M optical interface using the connection
cable.
Start enSuite and establish the communication link to the EK205-M ( section
communication link to the EK205-M via optical
Select "Navigation" on the left-hand side and then
of the navigation window.
At the top of the device window, open the path "Parameters" > "Interfaces" by clicking on
the relevant plus signs, and then select
In call acceptance windows 3 and 4, set start and end to the same value for each
one and send the changes using the
Then wait a minute for any communications activity in progress on the EK205-M to be safely stopped.
b) Sending the parameter file
• Send the following
parameter file to the
EK205-M (section
7.6 Send parameter
file, p. 51):
EK205_1v0. - RS232 (without Modem)
On the right, you can see
the navigation path to this
parameter file in enSuite for
software versions 1.0x of
the EK205-M.
instromet.com.
interface, p. 50)
"Parameterize
"Call time
windows".
"Transfer changes to
7.1 Setting up the
online" at the bottom
device" icon.
73