Installation, connection and commissioning
6.3.1.14
Close and secure calibration key
After making all settings subject to calibration requirements, close the calibration lock. Press the
key described in section 6.3.1.1 again.
The calibration lock is calibrated with an adhesive label as described in section 6.3.2.2.

6.3.2 Sealing

6.3.2.1 External view
6.3.2.2 Internal view
4
1
Fig. 17(a)
1
Fig. 16
2
3
5
Fig. 18
1 Optional user protection: Lid sealing
using wiring seals through sealing
plugs.
1
1 Sealing point for securing the
calibration switch.
2 Sealing points for securing the plate
covers.
3 Sealing points for securing the terminal
covers of the inputs and outputs.
4 Sealing points for securing the basic
device battery cover (if available).
5 Sealing points for securing the LCD
plate cover.
6
Sealing points for securing the modem
plate (bracket).
7
Sealing points for securing the modem
assembly.
6
7
Fig. 17(b)