Commissioning; Setting The Operating Parameters - Honeywell EK205-M Operating Instructions Manual

Volume conversion device
Table of Contents

Advertisement

Installation, connection and commissioning
6.3

Commissioning

6.3.1 Setting the operating parameters

The setting of the operating parameters required can be carried out using the free enSuite
software, see chapter 1.1.2).
If the program is not available, the settings can be alternatively made using the device keyboard.
Before setting the operating parameters using the keyboard, read section 9 to
familiarize yourself with the operation of the device.
6.3.1.1 Opening the calibration lock
On circuit board in the housing floor, the calibration lock can be found in the form of a key that
can be secured with an adhesive label. This needs to be operated to change relevant values and
parameters.
Fig. 15
6.3.1.2 Setting parameters for the NF pulse transducer of the gas meter
If a low-frequency pulse transducer is connected in accordance with 6.2.1.1, you can set the input
mode and cp value as follows:
1. Setting the input mode:
Move the cursor to the "Serv" tab and via the following path to the value "Md.I1" input
mode:
Serv.
Inputs
Press the key combination ENTER  +   The value set flashes.
Repeatedly press one of the  or  arrow buttons until the text "Pulse input" flashes.
Press the ENTER  +  button to confirm the set value. It is possible to abort the entry
by pressing the key combination ESC  + .
2. Setting the cp value (pulse constant):
Move the cursor to cp value "cp.I1" in the same path
Press the key combination ENTER  +   One digit of the value flashes.
Move the cursor with the arrow buttons  or  to the digits and change these with the
arrow buttons  or .
Input 1
Md.I1
Position of the key for
opening the calibration
lock (top assemblies
removed for clarity).
41