MODEL 7561-PSDS
PORTABLE SENSOR DISPLAY – STRAIN BRIDGE
TABLE 25. RANGE 1 TO 6 SCREEN
Item
Type drop down
Points button
Initials header
Live Input label
Live Output label
Calibration Curve chart
Undo button
Import button
Export button
Description
Select the type of calibration to perform. Options are:
Gain & Offset - Calibration with a single gain and offset
Multi Point -
Multi point calibration using multiple
gain and offset segments
Polynomial -
Multi point calibration using polynomi-
als
Add or remove a new calibration point. Calibration
points can only be added one at a time.
(Not present on the Gain & Offset calibration type)
,
NOTE
The advanced re-entrant calibration functionality
allows you to add calibration points in any order. The
user can even come back to the calibration later to add
more points or remove them. For example, the user
may perform various points in compression then the
next day return to add more points in tension
Enter calibrator initials
Indicates the live value of the mV/V input
Shows the live value of the output using the current
calibration settings
This is a coarse view of the calibration points that may
help to spot erroneous calibration points
Allows the last performed step to be undone
Import a previously exported calibration data file
Export the current calibration to a file for later import
into the same or a different handheld
Model 7561-PSDS User Manual |
Issue A
008-0752-00
sps.honeywell.com/ast
| 62