Honeywell 7561-PSDS User Manual page 24

Portable sensor display
Table of Contents

Advertisement

MODEL 7561-PSDS
PORTABLE SENSOR DISPLAY – STRAIN BRIDGE
|-CALIBRATION u
|-CALIBRATION INFOu
|
|-USER CALIBRATION u
|-RANGE NAME u
| |-RANGE 1
| |-RANGE 2
| |-RANGE 3
| |-RANGE 4
| |-RANGE 5
| |-RANGE 6
|-SELECT RANGE u
| |
| |
| |-—RANGE 1
| |-—RANGE 2
| |-—RANGE 3
| |-—RANGE 4
| |-—RANGE 5
| |-—RANGE 6
|-LIVE CALIBRATION u
| |-SENSITIVITY
| | |
| | |-—7.5 mV/V
| | |-—15 mV/V
| | |-—30 mV/V
| | |-—60 mV/V
| | |-—120 mV/V
| | |-—240 mV/V
| | |-—480 mV/V
| |-UNIT
| | |-—kg
| | |-—kilograms
| |
| |-LOW/HIGH CALIBRATION
| | |
| | |-SET OUTPUT LOW
| | |
| | |-SET OUTPUT HIGH
| | |
| | |-CAPTURE mV/V LOW
| | |
| | |-CAPTURE mV/V HIGH
| | |
| |
| |
| |-APPLY———
|
|
This menu contains settings regarding Calibration
Shows information about the currently selected calibration range.
Information: Name, Mode, Units, Type and last calibration date
Only available when no active TEDS device is attached
Only available when no activeTEDS device is attached
Select a Calibration Range item to display and edit its text
description
Some Calibration Ranges may not be available
Only available when no active TEDS device is attached. This allows
you to select the current Calibration Range to use. Some ranges
may not be available
Select Calibration Range 1
Select Calibration Range 2
Select Calibration Range 3
Select Calibration Range 4
Select Calibration Range 5
Select Calibration Range 6
Only available when no active TEDS device is attached
Select the desired sensitivity. Select an input range that covers the
strain bridge being connected
Select the desired sensitivity
Select the units that the calibration is to be performed in
Select desired units by highlighting either the units symbol line or
the units description line and press OK. This will cycle through all
available units (example shown for kilograms)
Configure the low and high points to perform calibration. Only a
two-point calibration can be performed using the keypad
Enter a value in the units (selected above) as the low point of the
known two-point calibration
Enter a value in the units (selected above) as the high point of the
known two-point calibration
Capture the inputs for mV/V to represent the two outputs entered
above. If the user actually applies the known force to the input
when selecting these options, the detected mV/V will be displayed
ready to either select OK or to manually edit
Use all of the information captured above to perform the actual
calibration
Model 7561-PSDS User Manual |
Issue A
008-0752-00
sps.honeywell.com/ast
| 24