Honeywell 7561-PSDS User Manual page 59

Portable sensor display
Table of Contents

Advertisement

MODEL 7561-PSDS
PORTABLE SENSOR DISPLAY – STRAIN BRIDGE
TABLE 23. MEASUREMENT SCREEN
Item
Remove button
Gross live label
Minimum Limit entry
field
Value Too Low indicator
Maximum Limit entry
field
Value Too High indicator
Gross Min live label
Gross Max live label
Gross Delta live label
Scale Steady Band edit
field
Scale Steady Duration
edit field
Tare edit field
Tare Now button
Remove button
Net live label
Net Min live label
Net Max live label
Net Delta live label
Display Resolution entry
field
Zero Mask entry field
Reset Statistics button
Simulated Display
Display Value live label
,
NOTE
If a TEDS device is connected, this page will reflect the settings particular to that
device.
Description
Set the current System Zero value back to zero
This is the current Gross value
Shows the minimum limit below which the input is
deemed under range. The live value will be replaced
with a warning message
The input value is below the Minimum Limit
Shows the maximum limit above which the input is
deemed over range. The live value will be replaced with
a warning message
The input value is below the Maximum Limit
Shows the minimum value the Gross has reached since
last reset
Shows the maximum value the Gross has reached
since last reset
Shows the difference between the minimum and maxi-
mum value the Gross has reached since last reset
Enter the band within the display reading must remain
(in this range's calibrated units of measure), for the
duration specified below, before the Scale Steady flag
becomes active
The duration in seconds for the above Scale Steady
calculation
The current value of the Tare. This is subtracted from
the Gross to give Net
Use the current Gross value to set the value of Tare to
give a zero Net value
Set Tare value back to zero
This is the current Net value
Shows the minimum value the Net has reached since
last reset
Shows the maximum value the Net has reached since
last reset
Shows the difference between the minimum and maxi-
mum value the Net has reached since last reset
The current setting for the resolution of the display.
The resolution
Enter a value band within which the display will show zero
Reset the Net Delta, Gross Min, Gross Max, Net Min
and Net Max values
Simulated display value showing the effects of the Zero
Masking and Resolution settings
Model 7561-PSDS User Manual |
Issue A
008-0752-00
sps.honeywell.com/ast
| 59

Advertisement