Range 1 - 6 Tabs - Honeywell 7561-PSDS User Manual

Portable sensor display
Table of Contents

Advertisement

MODEL 7561-PSDS
PORTABLE SENSOR DISPLAY – STRAIN BRIDGE

Range 1 - 6 Tabs

A tab exists for each of the six calibration ranges. Click on a tab to view, edit or start a
calibration.
Figure 45. Range 1 to 6 Tab Screen
The common controls will be described here, but view the individual sections for
specifics regarding each calibration type.
,
NOTE
It is important to select Sensitivity before starting a calibration. If sensitivity is
changed after calibration this may result in incorrect measurement.
TABLE 25. RANGE 1 TO 6 SCREEN
Item
Name edit field
Mode label
Sensitivity drop down
Units drop down
Description
Shows the name assigned to this calibration range.
Click to enter a new name
Indicates the type of calibration used internally. By
default, the calibration modes are calibrated for 6-wire
connections to eliminate the effect of cable length on
the calibration. See connection section of the manual
where it explains how to wire up for a 4-wire connec-
tion to the strain bridge sensor
Choose a sensitivity range suitable for the connected
strain gauge. Choose the lowest range that exceeds
the sensitivity of the connected bridge:
±7.5 mV/V
±15 mV/V
±30 mV/V
±60 mV/V
±120 mV/V
±240 mV/V
±480 mV/V
Select the engineering units that this calibration range
will be calibrated in. Depending on the chosen units
here, this will limit the available conversion units to
those compatible with the selected units
Model 7561-PSDS User Manual |
Issue A
008-0752-00
sps.honeywell.com/ast
| 61