MODEL 7561-PSDS
PORTABLE SENSOR DISPLAY – STRAIN BRIDGE
TABLE 30. ALARM SCREEN
Item
Trigger drop down
Alarm Properties image
Alarm Flags indicator
Description
Select how the alarm responds to the source value
relative to the two limits:
Disabled -
The alarm remains off
Outside Limits - The alarm is active when the source
value is above the high level OR below
the low level
Inside Limits -
The alarm is active when the source
value is below the high level AND
above the low level
Above High -
The alarm is active when the source
value is above the high level
Below High -
The alarm is active when the source
value is below the high level
Above Low -
The alarm is active when the source
value is above the low level
Below Low -
The alarm is active when the source
value is below the low level
This indicates at what input level the alarm will become
active. If the input value (shown) falls within a red zone,
the alarm will activate. If it falls within a green zone, the
alarm will not activate.
This action is changed by selecting a different Trigger
above
This indicates which of the eight alarm flags are cur-
rently active
Model 7561-PSDS User Manual |
Issue A
008-0752-00
sps.honeywell.com/ast