MODEL 7561-PSDS
PORTABLE SENSOR DISPLAY – STRAIN BRIDGE
Overriding TEDS Units
Sometimes the units held in the TEDS configuration may be incorrect or the user
may want to use a different or custom unit. The device menu offers a method to use
the custom units mechanism instead of the units held in the TEDS device. See Menu
Structure.

Displayed Error Codes

If errors occur whilst connected to a TEDS device, an error code will be displayed on
the LCD screen.
Figure 2. Error Codes
TABLE 2. ERROR CODES
Bit
Error Description
0
1-wire EEPROM not supported
1
Read from 1-wire EEPROM failed
2
Write to 1-wire EEPROM failed
3
TEDS profile is too large to be written into the 1-wire EEPROM
4
TEDS decoding failed on a read
5
CRC in TEDS read failed
6
Read from TEDS failed
7
Error reading from standard TEDS template
8
Error reading from calibration TEDS template
9
TEDS contains too many calibration templates (>5)
10
Template 40 contains too many calibration points (>10)
11
Template 41 contains too many polynomial coefficients (>10)
12
Template 41 contains too many segments (>1)
13
Template 42 contains too many entries (>10)
Model 7561-PSDS User Manual |
Issue A
008-0752-00
sps.honeywell.com/ast
| 10