MODEL 7561-PSDS
PORTABLE SENSOR DISPLAY – STRAIN BRIDGE

Alarm

Figure 50. Alarm
The single, global alarm acts on the currently selected calibration range (or TEDS
device) and is solely based on the Source selected here.
TABLE 30. ALARM SCREEN
Item
Source drop-down
High Level edit field
Low Level edit field
Actions
Beep switch
Flash Display switch
Latched switch
Description
Select the required mode for the backlight to operate
in. Choices are:
Gross -
Base alarm on the Gross value
Net -
Base alarm on the Net value
Enter the value for the higher level alarm point
Enter the value for the lower level alarm point
Select on or off to determine whether the beeper
sounds when the alarm is active
Select on or off to determine whether the main display
flashes when the alarm is active
Select whether the alarm is only active while the source
value is outside of the selected trigger criteria. This is
Unlatched operation. Latched operation means that
the alarm stays active even when the source value is
within limits again and requires a manual reset (via
menu) to deactivate
Model 7561-PSDS User Manual |