MODEL 7561-PSDS
PORTABLE SENSOR DISPLAY – STRAIN BRIDGE
Where There is No Option to Connect to Strain Bridge Sensor Chassis
Cable shield should be connected to the PSDS Excitation -ve and remain
unconnected at the strain bridge sensor end.
Figure 23. No Option to Connect to Strain Bridge Sensor Chassis
Model 7561-PSDS User Manual |
Issue A
008-0752-00