7. Post-installation checks

7.1. Check sheet

Checkpoint
Checking the software version
Inputting the control PC serial
number
Identifying the imaging units
and setting the number of units
connected
Checking the operation unit
Inputting the image
compensation data
Checking the date and time
Checking the timing with the
X-ray generators
Inserting the backup floppy
disk and checking the backup
files
Checking the radiographic
condition table
Network connections
Preparations prior to X-raying
Checking the annotation
Performing the setting and
checks for image transmission
to external memory
Checking the image quality
2 Installation Manual
Checkpoint details
Check that the CXDI application, A/D board
firmware and capture firmware versions all match.
Input the serial number of the control PC to be used.
Set the number of sensors to be used.
Register the serial numbers of the sensors in the
terminal.
Set the contrast.
Set the brightness.
Set the gamma correction.
Input the image compensation data.
Check the serial numbers of the sensors.
Perform the settings for multiple sensors are used.
Set the date.
Set the time.
Set the time zone.
Adjust the pre-delay and post-delay.
Delete the X-rayed image obtained at the
adjustment stage.
Delete the body part buttons used for adjustment.
Perform the settings for each of the X-ray tubes
when more than one X-ray tube is used.
Create the backup files in floppy drive by
re-starting.
(Cannot be used in automobile)
kV
mA
msec or mAs
IP address
Subnet mask
Default gateway
Perform calibration.
Perform self-diagnosis.
Checking that the setting have been made in
accordance
Checking that the setting have been made in
accordance
Use SMPTE pattern to check the density on a linear
chart.