4.3.17. Tool Modes

(1) Purpose
The tool modes (startup options) are intended in order to check operation, and are used to
launch the CXDI application on the control PC by itself, and to display items that are not
normally displayed.
(2) Preparation
1) Connect the keyboard and the mouse to the control PC.
2) Delete the "ccrstart.bat" file from Startup.
3) Disconnect the imaging unit from the control PC.
(3) Notes
Perform the following operation before using the /np mode. Especially, be sure to back up
the exposure mode names and the customized settings before the operation.
1) When using /np with the same settings as that of the connected sensor unit
BodyPart**.ini file can be used as it is.
Example
Sensor1
Sensor2
2) When using /np with different settings from that of the connected sensor unit
Move the BodyPart**.ini file in the BodyParts folder to the desktop, etc. However, do not
move the Reference folder. If the BodyPart**.ini file is left in the BodyParts folder,
system will not be able to be started, as the sensor type of the BodyPart and the settings
do not match.
2 Installation Manual
Connection
Table
Stand
- 99 -
/np setting
Table
Stand