6) The following menu appears on the Hyper Terminal screen.
0: Send sensor commands
1: Debug commands
2: Memory
3: Log
4:
5:
6:
7:
8:
9: Misc
(4) Sensor Serial No. Check
1) On the menu screen explained in (3) – 6) above, select "0: Send sensor commands".
2) When "ch idx =" appears, type the channel number of the sensor you want to check, "1"
if the channel number is 1.
3) When "Cmd str=" appears, enter the command "//osvr".
* Make sure to type the correct command.
4) The sensor information is displayed. Check that the sensor serial number written on the
A/D board is the same as the image data name file (********.dp) entered at
"Installation of LANMIT Image Correction Data".
For CXDI-11/12:
CH1: $$OS returned 0000
CH1: $$VR XXXX XXXX XXXX 0100 0017
CH1: $$CXDI-11 sensor main code
For CXDI-22
CH1: $$OS returned 0000
CH1: $$VR XXXX XXXX XXXX 0400 0019
CH1: $$CXDI-22 AD main code
For CXDI-31
CH1: $$OS returned 0000
CH1: $$VR XXXX XXXX XXXX 0500 0002 XXXX XXXX
CH1: $$CXDI-31 DEB AD main code
5) If the sensor serial number of the A/D board in step 4) is correct, close HyperTerminal,
and then perform the exit procedure in step (5) –9).
If the sensor serial number is different, enter the sensor serial number here referring to
step (5) below.