4.3.9. TABLE SETUP Settings
(1) Purpose
Adjust the CXDI operation unit's TABLE SETUP to match the exposure conditions (X-ray
tube voltage, X-ray tube current, msec or mAs value) of the X-ray generator.
(2) Procedure
1) Boot the CXDI system.
2) Open the TABLE SETUP Change window from the Normal Exposure window.
System
SETUP MENU
3) Select the tabs to be changed and change the X-ray tube voltage, X-ray tube current, and
msec or mAs value data to match the exposure conditions of the X-ray generator.
* See the operation manual for the details of settings.
4) After finishing the changes, return to the Normal Exposure window and check that the
TABLE SETUP has been changed.
2 Installation Manual
SYS. SETUP
TABLE SETUP
- 44 -