5) Measure the densities of the 11 locations (0% to 100%) of test image grayscale on the film
or on the monitor. [Fig. 3]
* Measure the SMPTE image density on the film is measured using a densitometer.
Measure the SMPTE image density on the high-definition monitor using the gradation
analysis software.
* The data for the SMPTE test image grayscale transferred by the CXDI are the values for
the maximum density (3.20 in the case of the MLP 190) in the printer settings which have
been changed in 11 uniform steps.
2 Installation Manual
[Fig. 3]
- 80 -