7.6 Post-installation checks

1) Check sheet
Check item
Detector check
Access point settings
Preparations for imaging
CXDI Control Software NE
settings
X-ray imaging
Confirmation of image transfer
to RIS, PACS and the printer
2. Installation procedures
What is checked
There must be no scratch marks, etc. on the exterior
of the detector.
Establish the settings using the access point setting
tool.
Turn on the power of the detector. Check that the
detector LEDs light.
Start the LNK Controller.
The detector must be registered by infrared
communication.
Communication with the detector must be verified.
The IP addresses of the detector must differ between
wireless and wired communication.
Check the remaining battery charge using GUI.
Conduct the discrete operation checks.
Establish the generator settings.
Establish the various settings.
No subject
Calibration
Self-diagnosis
Performance test
Data must be transferred by wireless
communication.
・ An image is transferred.
・ Confirmation of transfer data
Transfer test data to RIS, PACS and the printer and
the like, then check that there are no differences in
the data.
- 45 -