(3) Startup method
1) Start up Windows NT.
2) Launch "ccrstart.bat" which is located in "D:\ccr"
*When the CXDI application is launched, the message Sensor not connected
appears. Click [OK].
3) When the exposure screen appears on the operation unit, press [Alt] + [Tab] on the
keyboard to enter debugging mode.
4) When Welcome to CCR appears, select "1. Set-Up...".
5) When Setting Mode (0: Normal, 1: Expert) [0=0x0]: appears, select "1: Expert."
6) When CCR SETUP MENU appears, select "7 Scan Sensor Setup".
7) When Capture Device Configuration Table appears, enter dummy serial numbers for the
items indicated below.
The dummy serial numbers are as follows:
Upright stand
Patient table
Universal
Cassette
* Be sure to write down the correct sensor serial numbers before entering the
dummy serial numbers.
For example, when the upright stand model is connected to Sensor 1, and the table model
is connected to Sensor 2:
Enter "0000199" for "A/D Board Serial Number for Sensor ID#1"
Enter "00002009" for "A/D Board Serial Number for Sensor ID#2"
The CXDI application can now be launched on the control PC with the same conditions
in effect as if an imaging unit were connected.
2 Installation Manual
Serial Number
Serial Number
Serial Number
Serial Number
- 100 -
00000199
00002009
0000123
0000302