(5) -I
-I
A-RELEASE-RES is ignored.
Meaning
* This parameter is used simply as "-I"
Description * This is used when the error message "30012 Peer aborted Association (or never
connected)" occurseven though the DICOM data transfer was successful.
This is used based on the connected storage devices.
(6) -d FAC
-d FAC
This parameter dumps a specific facility log.
(DCM/DUL/SRV)
Meaning
* This parameter is used simply as "-d"
* This parameter is used to make the transfer software display the debugging
character string on the console.
Description * This parameter does not affect DICOM data transfer.
* This parameter outputs the CXDI log based on Windows NT.
(7) -v
-v
This parameter dumps the transfer log.
Meaning
* This parameter is used simply as "-v".
* DUL and SRV are dumped.
* This parameter is used to make the transfer software display the debugging
character string on the console.
Description * This parameter does not affect DICOM data transfer.
(8) -jn
-jn
This is the time to take timeout.
Meaning
* Sets the time to take timeout in seconds.
Description * This parameter is to be changed when taking timeout.
(9) -k
-k
level = 0
1
2
Meaning
* A variety of specifications have since been needed in conjunction with DICOM
modality LUT support.
Description
* If DICOM Modality LUT OD is enabled, set appropriate options to suit each output
destination.
Argument : Data necessary for executing a function, subroutine, procedure, or other operation.
Arguments are assigned to functions and subroutines when executed. For example,
the argument in f (χ) is χ.
PDU :
Protocol data unit
The types of PDUs include get-request, get-next-request, get-response, set-request,
nd trap.
Note
For details about the -v parameter, see "Checking the Error Log".