5) As the step (3)-6), create a graph based on the data measured in step 4), and make sure that
the data from the minimum density to the maximum density nearly form a straight line.
3.50
3.00
2.50
2.00
1.50
1.00
0.50
0.00
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31
Ideal Density Data for Linear Line
Step
1
2
3.2 D
0
0.10
3.1 D
0
0.10
12 Bit
0
132
8 Bit
0
Step
12
13
3.2 D
1.14
1.24
3.1 D
1.10
1.20
12 Bit
1453
1585
8 Bit
90
98
Step
23
24
3.2 D
2.27
2.37
3.1 D
2.20
2.30
12 Bit
2906
3038
8 Bit
180
188
2 Installation Manual
Step
3
4
5
0.21
0.31
0.41
0.20
0.30
0.40
264
396
528
8
16
24
14
15
16
1.34
1.45
1.55
1.30
1.40
1.50
1717
1849
1981
106
114
122
25
26
27
2.48
2.58
2.68
2.40
2.50
2.60
3170
3302
3434
196
205
213
- 84 -
6
7
0.52
0.62
0.72
0.50
0.60
0.70
660
792
32
41
49
17
18
19
1.65
1.75
1.86
1.60
1.70
1.80
2113
2245
2377
131
139
28
29
30
2.79
2.89
2.99
2.70
2.80
2.90
3566
3698
3830
221
229
Max setting
density
Film base
density
8
9
10
0.83
0.93
0.80
0.90
924
1056
1188
57
65
74
20
21
1.96
2.06
1.90
2.00
2509
2641
147
155
164
31
32
3.10
3.20
3.00
3.10
3962
4095
238
246
255
11
1.03
1.00
1320
82
22
2.17
2.10
2774
172