Slg 700 Ff Level Transmitter Configuration; Importing The Slg 700 Ff Device Description (Dd) Files; Importing The Dd To Experion Pks - Honeywell SLG 700 Option Manual

Smartline level transmitter guided wave radar foundation fieldbus
Hide thumbs Also See for SLG 700:
Table of Contents

Advertisement

3. SLG 700 FF Level Transmitter Configuration

3.1 Importing the SLG 700 FF Device Description (DD) files

Importing the DD to Experion PKS

ATTENTION
Experion release compatibility
The steps in the following procedure are specific to Experion only.
Step
From the Control Builder main screen, click Fieldbus Device Description Import
1
OR
Select File > New > Type >Fieldbus Device
You can Import the DD using one of the following steps:
Choose Browse
2
Select the required folder, and click OK.
Select the DD from the Device List, and click OK.
The following dialog box appears,
3
Click OK.
Rev.5
Experion Release
DD Compatibility
431.1
Yes
430.3
Yes
410.7
Yes
to locate the folder where you have stored the DD file.