Parameter
PARAM_INDEX
UNIT_TYPES
CUSTOM_UNIT
CUSTOM_TAG
DISPLAY_TEMPLATE
DECIMALS
PV_LOLIM
PV_HILIM
Page 68
FOUNDATION Fieldbus Option User's Manual
Parameter selection for screen process variable. Parameters need
to be chosen based on Block type. The PARAM_INDEX is present in
all the eight screens: PARAM_INDEX_1, PARAM_INDEX_2,
PARAM_INDEX_3, PARAM_INDEX_4, PARAM_INDEX_5,
PARAM_INDEX_6, PARAM_INDEX_7 and, PARAM_INDEX_8.
Unit selection for screen process variable. Appropriate units need to
be selected based on the configured parameter. If desired units are
not present, 'custom' may be selected. The UNIT_TYPES is present
in all the eight screens: UNIT_TYPES_1, UNIT_TYPES_2,
UNIT_TYPES_3, UNIT_TYPES_4, UNIT_TYPES_5,
UNIT_TYPES_6, UNIT_TYPES_7 and, UNIT_TYPES_8.
Character string to represent custom units. This value is used when
Unit type of 'custom' is selected. Size: 8 Characters. The
CUSTOM_UNIT is present in all the eight screens:
CUSTOM_UNIT_1, CUSTOM_UNIT_2, CUSTOM_UNIT_3,
CUSTOM_UNIT_4, CUSTOM_UNIT_5, CUSTOM_UNIT_6,
CUSTOM_UNIT_7 and CUSTOM_UNIT_7.
Tag to be displayed for the screen. Length: 14 Characters. The
CUSTOM_TAG is present in all the eight screens:
CUSTOM_TAG_1, CUSTOM_TAG_2, CUSTOM_TAG_3,
CUSTOM_TAG_4, CUSTOM_TAG_5, CUSTOM_TAG_6,
CUSTOM_TAG_7 and CUSTOM_TAG_8.
Represents the display screen template. Possible Values:
a.
PV : PV value is displayed
b.
PV and Trend : PV value followed by a Trend is shown on the
display
c.
PV and Bar Graph : PV value followed by a Bargraph is shown
on the display
d.
None: Screen will not be seen.
The DISPLAY_TEMPLATE is present in all the eight screens:
DISPLAY_TEMPLATE_1, DISPLAY_TEMPLATE_2,
DISPLAY_TEMPLATE_3, DISPLAY_TEMPLATE_4,
DISPLAY_TEMPLATE_5, DISPLAY_TEMPLATE_6,
DISPLAY_TEMPLATE_7 and DISPLAY_TEMPLATE_8.
Number of digits to display after the decimal point. Range: 0 - 3.
DECIMALS are present in all the eight screens: DECIMALS_1,
DECIMALS_2, DECIMALS_3, DECIMALS_4, DECIMALS_5,
DECIMALS_6, DECIMALS_7 and DECIMALS_8.
Display Low Limit (Trend, Bar, Custom PV scaling, usually equal to
LRV). The PV_LOLIM is present in all the eight screens:
PV_LOLIM_1, PV_LOLIM_2, PV_LOLIM_3, PV_LOLIM_4,
PV_LOLIM_5, PV_LOLIM_6, PV_LOLIM_7 and PV_LOLIM_8.
Display High Limit (Trend, Bar, Custom PV scaling, usually equal to
URV). The PV_HILIM is present in all the eight screens:
PV_HILIM_1, PV_HILIM_2, PV_HILIM_3, PV_HILIM_4,
PV_HILIM_5, PV_HILIM_6, PV_HILIM_7 and PV_HILIM_8.